TCAV-L Averaging Soil Thermocouple Probe

Overview

The TCAV-L typically provides the average temperature of the top 6 to 8 cm of soil for energy-balance in flux systems. It parallels four thermocouples together into one 24 AWG wire. Each member of a thermocouple pair can then be buried at a different depth. The two pairs are separated at a distance of up to 1 m.

Read More

Benefits and Features

  • Used in conjunction with soil heat-flux plates and soil-moisture probes to compute total soil heat flux

Images


Specifications

Type Chromel-Constantan
Typical Output 60 μV/°C
Accuracy Refer to the Thermocouple Measurement section in the data logger manual.
Weight 0.45 kg (1 lb) with 15.24-m (50-ft) cable

Compatibility

Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.

Data Loggers

Compatible Note
21X (retired)
CR10 (retired)
CR1000
CR1000X
CR10X (retired)
CR200X (retired)
CR206X (retired)
CR23X (retired)
CR3000
CR500 (retired)
CR5000 (retired)
CR510 (retired)
CR6
CR800
CR850
CR9000 (retired)
CR9000X

Additional Compatibility Information

Data Logger Considerations

One differential analog input channel per probe is required.

Reference Temperature Measurement

A reference temperature measurement  is required. Options for measuring the reference temperature include:

  • Thermistor built into the CR800, CR850, CR1000, CR3000, or CR5000 wiring panel
  • PRT built into the wiring panel of the CR9050 or
  • CR9051E input module for the CR9000X Measurement and Control Datalogger
  • PRT built into the wiring panel of the CR723T input card for the CR7 Measurement and Control Datalogger
  • CR10XTCR thermistor that connects to the CR10X wiring panel

FAQs for

Number of FAQs related to TCAV-L: 4

Expand AllCollapse All

  1. Yes, as long as the TCAV-L is measured using a differential measurement instruction. Be sure to connect the shield wire to analog ground.

  2. The TCAV-L is constructed using a Type E thermocouple wire. The high-order NIST polynomials were fit over the temperature range of -270° to 1000°C. The temperature range of a TCAV-L on a data logger depends on the measurement range selected in the thermocouple measurement instruction. For example, using a CR3000 and the ±20 mV input range and a reference temperature at 20°C, the TCAV-L temperature range is -270° to 300°C. Refer to the measurement section in the appropriate data logger manual for a detailed discussion on thermocouple measurements.

  3. The TCAV-L measurement range, accuracy, and resolution are a function of the data logger used to measure the TCAV-L. Refer to the appropriate data logger manual measurement section for more information.

  4. The TCAV-L is a Type E thermocouple. Use a Type E (chromel –constantan) Thermocouple found in the Sensor | Temperature group.