HMP155A Temperature and Relative Humidity Probe
Accurate, Wide Temp Range
Higher-end sensor where higher accuracy is required
weather applications supported water applications supported energy applications supported gas flux and turbulence applications supported infrastructure applications supported soil applications supported

Overview

The HMP155A provides reliable relative humidity (RH) and temperature measurements for a wide range of applications. It uses a HUMICAP®180R capacitive thin film polymer sensor to measure RH over the 0 to 100% RH range. A PRT measures temperature over the -80° to +60°C range. This rugged, accurate temperature/RH probe is manufactured by Vaisala.

Manufacturer's note: In changing temperature conditions with high humidity, the probe temperature can lag behind the temperature of surrounding air and this can cause risk of condensation on the sensor. A wet probe cannot observe the actual humidity in the ambient air. If the condense water is contaminated, the life span of the probe may shorten and the calibration may change. 

Read More

Benefits and Features

  • Well-suited for long-term, unattended applications
  • Accurate and rugged
  • Mounts to a mast, crossarm, or user-supplied pole
  • Compatible with most Campbell Scientific data loggers

This product is available for repair and calibration

Images

Detailed Description

To reduce the current drain, power can be supplied to the HMP155A only during measurement when the sensor is connected to the data logger’s switched 12 V terminal. Data loggers that do not have a switched 12 V terminal, such as the CR510 or CR7, can use the SW12V switched 12 V device to switch power to the sensor only during measurement.

Compatibility

Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.

Data Loggers

Compatible Note
CR1000 (retired)
CR1000X (retired)
CR300 (retired)
CR3000 (retired)
CR310
CR350
CR6
CR800 (retired)
CR850 (retired)

Additional Compatibility Information

Sensor Mounts

When exposed to sunlight, the HMP155A must be housed in a 41005-5 or RAD14 14-plate naturally aspirated radiation shield. The 41005-5 and RAD14 attach to a crossarm, mast, or user-supplied pipe with a 2.5 to 5.3 cm (1.0 to 2.1 in.) outer diameter. The RAD14 uses a double-louvered design that offers improved sensor protection from driving rain, snow, and insect intrusion, It also has lower self-heating in bright sunlight combined with higher temperatures (> 24°C [~75°F])
and low wind speeds (< 2 m s-1 [~4.5 mph]), giving a better measurement.

Specifications

Electromagnetic Compatibility Complies with EMC standard EN61326-1 Electromagnetic
Filter Description Sintered PTFE
Housing Body Material PC
Housing Classification IP66
Voltage Output 0 to 1 Vdc
Average Current Consumption ≤ 3 mA (analog output mode)
Operating Voltage 7 to 28 Vdc
Settling Time 2 s (at power up)
Field-Replaceable Chip or Recalibrate Recalibrate
Tip Diameter 1.2 cm (0.5 in.)
Length 27.9 cm (11 in.)
Head Height 4 cm (1.6 in.)
Body Height 2.4 cm (0.9 in.)
Body Width 2.0 cm (0.8 in.)

Relative Humidity

Sensing Element HUMICAP 180R
Measurement Range 0 to 100% RH (non-condensing)
Response Time
  • The response time for the RH specification is for the HUMICAP 180R at 20°C in still air with sintered PTFE filter and a 0 to 75% RH step change.
  • 20 s (63% step change)
  • 60 s (90% step change)
Factory Calibration Uncertainty
  • The factory calibration uncertainty is defined as ±2 standard deviation limits. Uncertainty is at +20°C. Small variations are possible; see also the calibration certificate.
  • ±0.6% RH 0 to 40% RH
  • ±1.0% RH 40 to 97% RH
Accuracy
  • -NOTE- Accuracy specifications include non-linearity, hysteresis, and repeatability.
  • ±1% RH (at 15° to 25°C, 0 to 90% RH)
  • ±1.7% RH (at 15° to 25°C, 90 to 100% RH)
  • ±(1.4 + 0.032 × reading) % RH (at -60° to -40°C)
  • ±(1.2 + 0.012 × reading) % RH (at -40° to -20°C)
  • ±(1.0 + 0.008 × reading) % RH (at -20° to +40°C)
  • ± (1.2 + 0.012 × reading) % RH (at 40° to 60°C)

Air Temperature

Sensing Element PT 100 RTD 1/3 class B IEC 751
Measurement Range -80° to +60°C
Accuracy
  • ±(0.226 - 0.0028 x temperature)°C (-80° to +20°C)
  • ±(0.055 + 0.0057 x temperature)°C (+20° to +60°C)
Entire Temperature Range Refer to graph in probe manual.


FAQs for

Number of FAQs related to HMP155A: 3

Expand AllCollapse All

  1. Note the difference between calibration and a field check. Calibration cannot be done in the field, as it requires an experienced technician and specialized equipment.

    Field checks of measurements can be done to determine if the data make sense with the real-world conditions. Follow these steps to field check a sensor:

    1. Find a second sensor of the same type as the installed sensor whose data is in question. The second sensor will be used as a benchmark sensor and should be known to be accurate or recently calibrated.
    2. At the site, take readings using both sensors under the same conditions. The best practice is to measure both sensors side-by-side at the same time. Note that the sensors will never have the exact same measurement.
    3. Depending on the sensor model, if the difference in the readings of the installed and benchmark sensors is greater than the sum of the accuracies for both sensors, either return the installed sensor to Campbell Scientific for calibration or replace the appropriate chip.
      • The 107, 108, 109, 110PV-L, and BlackGlobe-L temperature sensors can be calibrated.
      • The HC2S3-L and HMP155A-L temperature and relative humidity sensors can be calibrated.
      • The CS215-L has a replaceable chip for temperature and relative humidity. For more information, refer to the “Maintenance and Calibration” section of the CS215 instruction manual.
      • The HMP60-L has a replaceable chip for relative humidity only. For more information, refer to the “Maintenance” section of the HMP60 instruction manual.
  2. Single-ended channels are typically used when it is desirable to use fewer channels and the cable length is shorter—less than 6.1 m (20 ft). Differential channels are preferable when the cable needs to be longer than 6.1 m (20 ft).