The quality metric is a unique diagnostic feature that indicates the quality of the pressure measurement returned by the sensor card (BVC10). It enables the user to monitor the health of their sensor to better schedule recalibrations.
Quality Metric Range | Description | Comments |
> 6 |
Sensor freshly calibrated |
|
5 to 5.99 |
Sensor working well |
|
4 to 4.99 |
In two to three months, this sensor will need to be recalibrated. |
Within three months, send the BaroVue 10 to the factory for recalibration. |
0 to 3.99 |
Recalibrate the sensor now. |
As soon as possible, send the BaroVue 10 to the factory for recalibration. |