Hydrolab DataSonde® 4X Multiprobe (includes Central Cleaning Unit)

Click picture for full size image.
Manual, Version Date: April, 1998  Manual

White Paper, Version Date: April, 2004  White Paper

Software and Drivers
Technical Notes

General Description

The DS4X offers Hydrolab's Superior Sensor Technology on a platform that will dramatically extend your DataSonde's deployment life!

Benefits

  • Ideal for "X-tended" deployments in environments where fouling and sediment are abundant
  • Further improved long-term value due to reduced frequency of sensor maintenance
  • Central cleaning system wipes away fouling on DO, pH, ISE's, Chlorophyll, and Turbidity
  • One central motor cleans the entire suite of sensors – minimizes power consumption!
  • Hydrolab's brush design features robust fibers that won't separate over time

Parameter Specifications

Parameter Range Accuracy Resolution Available Instrument *
Temperature -5°  to 50°C ±0.10°C 0.01°C DS4X
Specific Conductance 0 to 100 mS/cm ±1% of reading;
±0.001 mS/cm
4 digits DS4X
pH 0 to 14 units ±0.2 units 0.01 units DS4X
Dissolved Oxygen 0 to 50 mg/L ±0.2 mg/L at <= 20mg/L
±0.6 mg/L at > 20mg/L
0.01 mg/L DS4X
ORP -999 to 999 mV ±20 mV 1 mV DS4X
Depth Vented Level 0 to 10 m ±0.003 m 0.001 m DS4X
  0 to 25 m ±0.05 m 0.01 m DS4X
  0 to 100 m ±0.05 m 0.01 m DS4X
  0 to 200 m ±0.1 m 0.1 m DS4X
Salinity 0 to 70 ppt ±0.2 ppt 0.01 ppt DS4X
Self-Cleaning Turbidity 0 to 3000 NTU ±1%, up to 100 NTU
±3%, 100-400 NTU
±5%, 400-3000 NTU
0.1, up to 400 NTU
1.0, 400-3000 NTU
DS4X
Ammonium/
Ammonia
0 to 100 mg/L-N Greater of ±5 of reading or ±2  mg/L-N (typical) 0.01 mg/L-N DS4X
Nitrate 0 to 100 mg/l-N Greater of ±5 of reading or ±2 mg/L-N (typical) 0.01 mg/L-N DS4X
Chloride 0.5 to 18,000 mg/L Greater of ±5 of reading or ±2 mg/L (typical) 4 digits DS4X
Total Dissolved Gas 400 to 1300 mmHg ±0.1% of span 1.0 mmHg DS4X
Ambient Light 0 to 10,000 µmol s-1m-2 ±5% of reading or ±1 µmol
s-1m-2
1 µmol s-1m-2 DS4X
Chlorophyll a 0 to 500 µg/L
0 to 50 µg/L
0 to 5 µg/L
  0.01 µg/L DS4X
Barometric Pressure 500 to 850 mmHg ±10 mmHg 0.1 mmHg S
Global Positioning System -90 to 90° Latitude
-180 to 180° Latitude
25 m CEP (50%) without SA and DGPS
2m CEP (50%) with DGPS
0.1" S
* DS4X = DataSonde 4X, S = Surveyor 4a

Physical Specifications

  Length Diameter Weight
with weighted sensor guard 23 in
58.4 cm
3.5 in
8.9 cm
7.4 lbs
3.35 kg
 
Maximum Immersion Depth: 225 meters
Operating Temperature Range: -5° to 50°C

Download Adobe Acrobat Reader



Home | What's New | Products | Services | Applications | Courses & Shows | Museum | Literature | Privacy | About Us
Contact Us | Recruitment | Search | Links | Conversions | Download | Site Map

Questions? Please refer to the email address in this graphic if the adjacent link does not work.   Phone: (780) 454 2505
© 2004 Campbell Scientific (Canada) Corp.  
Top of Page