Product Information - CS511-L (Discontinued)

Print this page

Sensorex Dissolved Oxygen Probe

Click picture for full size image.
Model Cable Length
CS511-L Custom

Manual, Version Date: January 31, 2001  Manual

Features

  • Measurement range 0 to 20 mg l-1 dissolved oxygen
  • Temperature Compensation from +40° to +104°C
  • Dimensions are 2.2 in diameter x 3.5 in height
  • Requires 2.5 in submersion depth; 2 in s-1 minimum water flow across membrane (agitator available)
  • Manufactured by Sensorex

General Description

The CS511-L is a rugged, low-maintenance sensor that is manufactured by Sensorex. It consists of a self-polarizing galvanic cell that generates a millivolt signal proportional to the amount of oxygen present in the measured medium (typically water). Oxygen diffuses through its membrane onto a cathode which produces a chemical reaction. An electrical current is produced by this reaction which is converted from microamps to millivolts by an in-line resistor. An in-line thermistor provides automatic temperature compensation.

Options

The PT4-L agitator is for applications where the water velocity is less than 1 inch per second (2.5 cm s-1). The PT4-L agitator consists of 12 Vdc solenoid and toggle mechanism that sweep a brush across the probe membrane. A PFS timer controls the sweeping frequency to keep the probe clean and provide water movement for accurate readings.

Service kits are available for easy replacement of sensor components in the field; contact our office for details.

Specifications

Range: 0 to 20 mg l-1 dissolved oxygen, 0 to 200% sat
Output: 1.65 mV ±0.45 mV per mg l-1
Accuracy: ±2%
Response Time: After equilibration, 1 minute for 95% of final value
Minimum Water Velocity: 2 in s-1 (5 cm s-1) across membrane
Minimum Submersion Depth: 2.5 in (6.35 cm)
Operating Temperature Range: 0° to 50°C (32° to 122°F)
Pressure: 0 - 100 psig
Temperature Compensation: Automatic from 4° to 40°C (40° to 104°F)
Electrode Materials: Ag cathode/Zn anode
Probe Electrolyte: NaCl
Dimensions: Diameter: 5.59 cm (2.2 in)
Probe Height: 8.89 cm (3.5 in)
Weight: 0.50 kg (1.1 lbs)
(includes 10 feet of 2-conductor, 16-gage cable)
Power Requirements for optional PT4 Agitator: 12 Vdc, 0.9 A
Datalogger Compatibility: CR510, CR10X, CR23X, CR5000, CR7, CR9000
Call for compatibility with older dataloggers.
Download Adobe Acrobat Reader



Home | What's New | Products | Services | Applications | Courses & Shows | Museum | Literature | Privacy | About Us
Contact Us | Recruitment | Search | Links | Conversions | Download | Site Map

Questions? Please refer to the email address in this graphic if the adjacent link does not work.   Phone: (780) 454 2505
© 2001 Campbell Scientific (Canada) Corp.  
Top of Page